| articles:nist_traceability [2025/11/18 20:28] – [An "unbroken chain of calibrations" to what?] rrandall | articles:nist_traceability [2025/11/18 20:33] (current) – [An "unbroken chain of calibrations" to what?] rrandall | ||
|---|---|---|---|
| Line 54: | Line 54: | ||
| (b) (1) Calibration standards. Calibration standards used for inspection, measuring, and test equipment shall be __traceable to national or international standards__. If national or international standards are not practical or available, the manufacturer shall use an independent reproducible standard. If no applicable standard exists, the manufacturer shall establish and maintain an in-house standard.</ | (b) (1) Calibration standards. Calibration standards used for inspection, measuring, and test equipment shall be __traceable to national or international standards__. If national or international standards are not practical or available, the manufacturer shall use an independent reproducible standard. If no applicable standard exists, the manufacturer shall establish and maintain an in-house standard.</ | ||
| - | <WRAP info center round box 80%>On February 2, 2024, the U.S. FDA issued an [[https:// | + | <WRAP info center round box 80%>On February 2, 2024, the U.S. FDA issued an [[https:// |
| Similar antiquated language appears in [[http:// | Similar antiquated language appears in [[http:// | ||