 ====== What are Accuracy Ratios ====== ====== What are Accuracy Ratios ======
 +{{keywords>accuracy ratio}}
 Many users overlook the requirement in both ISO 9001 & AS9100D that states:  Many users overlook the requirement in both ISO 9001 & AS9100D that states: 
 <blockquote>7.1.5 Monitoring and measuring resources \\ 7.1.5.1 General \\  <blockquote>7.1.5 Monitoring and measuring resources \\ 7.1.5.1 General \\ 
 //The organization shall determine and provide the resources needed to ensure valid and reliable results when monitoring or measuring is used to verify the conformity of products and services to requirements.//</blockquote> //The organization shall determine and provide the resources needed to ensure valid and reliable results when monitoring or measuring is used to verify the conformity of products and services to requirements.//</blockquote>
  
-To "//ensure valid and reliable results//" the company must provide personnel with "monitoring or measuring" instruments (resources) that possess sufficient accuracy, range & resolution.  A "key" point here is the instruments possess a sufficient "Accuracy Ratio".+To "//ensure valid and reliable results//" the company must provide personnel with "monitoring or measuring" instruments (resources) that possess sufficient accuracy, range & resolution.  A "key" point here is that the instruments possess a sufficient "Accuracy Ratio".
  
 An "Accuracy Ratio" is defined in [[https://www.sae.org/standards/content/as13003/|SAE AS13003, "Measurement Systems Analysis Requirements for the Aero Engine Supply Chain"]], as: An "Accuracy Ratio" is defined in [[https://www.sae.org/standards/content/as13003/|SAE AS13003, "Measurement Systems Analysis Requirements for the Aero Engine Supply Chain"]], as:
Line 16: Line 16:
 <WRAP centeralign>**Dimension tolerance being measured ÷ Accuracy of measuring device used = Accuracy Ratio** </WRAP> <WRAP centeralign>**Dimension tolerance being measured ÷ Accuracy of measuring device used = Accuracy Ratio** </WRAP>
 <WRAP clear></WRAP> <WRAP clear></WRAP>
 +{{ :articles:micrometer-pngrepo-com.png?direct&60|Source: https://www.pngrepo.com/svg/234351/micrometer}}
 Using another example, let's assume that you have a micrometer with an accuracy of ±0.0001". And you're using it to measure a dimensional feature with a tolerance of ±0.005". Divide 0.005 by 0.0001 to obtain your accuracy ratio. In this scenario, you'd have an accuracy ratio of 50:1. Using another example, let's assume that you have a micrometer with an accuracy of ±0.0001". And you're using it to measure a dimensional feature with a tolerance of ±0.005". Divide 0.005 by 0.0001 to obtain your accuracy ratio. In this scenario, you'd have an accuracy ratio of 50:1.
  
-Management of accuracy ratio is a “risk control”The greater the accuracy ratio, the smaller the likelihood that an Out-of-Tolerance condition will impact the feature measured. Conversely, the lower the accuracy ratio, the greater the likelihood that an Out-of-Tolerance condition will impact the feature measured (and the part).+===== Management of accuracy ratio is a “risk control” ===== 
 + 
 +The greater the accuracy ratio, the smaller the likelihood that an Out-of-Tolerance condition will impact the feature measured. Conversely, the lower the accuracy ratio, the greater the likelihood that an Out-of-Tolerance condition will impact the feature measured (and the part). 
 + 
 +Calibration Labs typically apply a minimum 4:1 "Test Accuracy Ratio" (TAR) rule using a similar concept. Generally speaking, 4:1 is the lowest //accuracy ratio// that should be maintained in order to "//ensure valid and reliable results//" (readings). 
 + 
 +Let's suppose that a measuring device is found Out-of-Tolerance. A measuring device that is found Out-of-Tolerance ≥200% over its stated accuracy is considered "//significantly//" Out-of-Tolerance (SOOT). Assume that a dimensional feature with a tolerance of ±0.001" was measured using an instrument accurate to ±0.0001". If this measuring instrument was found to be 200% Out-of-Tolerance (a SOOT condition), multiply the instrument accuracy (0.0001") by 2. Then determine your new accuracy ratio using the calculation of 0.001 ÷ 0.0002, which results in a new accuracy ratio of 5:1. Even with the measuring device "significantly" Out-of-Tolerance, the accuracy ratio (in this instance) is still above 4:1. So there is a high degree of probability that conforming product was still delivered to the customer. 
 + 
 +However, if the dimensional feature had a tolerance of ±0.001", and was measured using an instrument accurate to ±0.00015". Multiply that instrument accuracy (0.00015") by 2. Then determine your new accuracy ratio using the calculation of 0.001 ÷ 0.0003, which results in a new accuracy ratio of 3.3:1. Since this accuracy ratio is below 4:1, then this would reveal an increased likelihood that nonconforming product had been delivered to the customer. 
 + 
 +===== Out-of-Tolerance Impact Analysis ===== 
 + 
 +Many users also overlook the requirement in both ISO 9001 & AS9100D that states:  
 +<blockquote>7.1.5.2 Measurement traceability \\  
 +//The organization shall determine if the validity of previous measurement results has been adversely affected when measuring equipment is found to be unfit for its intended purpose, and shall take appropriate action as necessary.//</blockquote>
  
-Calibration Labs typically apply a minimum 4:1 "Test Accuracy Ratio(TAR) rule using a similar concept. Generally speaking, 4:1 is the lowest accuracy ratio that should be maintained in order to achieve reasonable confidence that conforming parts are being delivered.+The best, and easiest way to "//determine if the validity of previous measurement results//" is to, as described above, calculate the "accuracy ratio" based on the results ("As Found" condition) reported by the Calibration Lab.
  
-Let's suppose that measuring device is found Out-of-Tolerance. A measuring device that is found Out-of-Tolerance ≥200% over its stated accuracy is considered "//significantly//Out-of-Tolerance (SOOT)Assume that dimensional feature with a tolerance of ±0.001" was measured using an instrument accurate to ±0.0001". If this measuring instrument was found to be 200% Out-of-Tolerance (a SOOT condition), multiply the instrument accuracy (0.0001") by 2. Then determine your new accuracy ratio using the calculation of 0.001 ÷ 0.0002, which results in a new accuracy ratio of 5:1. Even with the measuring device "significantlyOut-of-Tolerance, the accuracy ratio (in this instance) is still above 4:1. So this would be justification for identifying the nonconformity as a minor.+If the diminished "Accuracy Ratio" remained above 4:1, then there is reasonable degree of confidence that the OOT (Out-of-Tolerance) condition did not "adverselyaffect the delivered product or serviceHowever, if the diminished "Accuracy Ratio" fell below 4:1, then there is reasonable degree of uncertainty in whether the OOT (Out-of-Tolerancecondition "adversely" affected the delivered product or serviceAnd the lower the "Accuracy Ratio" (<4:1), the greater the probability that the delivered product or service had been "adversely" affected.
  
-However, if the dimensional feature had a tolerance of ±0.001", and was measured using an instrument accurate to ±0.00015". Multiply that instrument accuracy (0.00015") by 2. Then determine your new accuracy ratio using the calculation of 0.001 ÷ 0.0003, which results in a new accuracy ratio of 3.3:1. Since this accuracy ratio is below 4:1, then this could be justification for a major nonconformity.