Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
articles:auditing_7.1.5 [2024/07/09 14:54] – [6 - Using a Precision Micrometer to calibrate Gage Blocks] rrandallarticles:auditing_7.1.5 [2025/05/02 01:57] (current) – [11 - Counting & Weighing Scales marked "For Reference Only"] rrandall
Line 32: Line 32:
   * calibration or verification method, and    * calibration or verification method, and 
   * acceptance criteria.   * acceptance criteria.
 +
 +[[articles:auditing_7.1.5#Counting & Weighing Scales marked "For Reference Only"|Scenario #11:]] While auditing a Distributor, you observe their weighing scales, counting scales and length meters (for measuring the length of cable & wire), all identified as “For Reference Only”.
  
 The company explained that, since all of their calibrations are outsourced, this information can be found in the calibration certificates provided by the external calibration lab. The company explained that, since all of their calibrations are outsourced, this information can be found in the calibration certificates provided by the external calibration lab.
Line 324: Line 326:
 <blockquote>//The 9100:2009-series verbiage require a calibration register and the definition of processes for calibration/verification (including equipment type, ID, frequency, methods and acceptance criteria), but didn’t seem to require them to be one in the same. The 9100:2016 standard appears to mandate these definitions be incorporated into the register itself, as opposed to just being defined. Is this required to be taken literally that the register is required to have this information is absolute?//</blockquote> <blockquote>//The 9100:2009-series verbiage require a calibration register and the definition of processes for calibration/verification (including equipment type, ID, frequency, methods and acceptance criteria), but didn’t seem to require them to be one in the same. The 9100:2016 standard appears to mandate these definitions be incorporated into the register itself, as opposed to just being defined. Is this required to be taken literally that the register is required to have this information is absolute?//</blockquote>
  
-The [[https://www.sae.org/iaqg/projects/9100-2016_clarification_table.pdf|IAQG "Official" AS 9100:2016 Series Clarifications]] contains the following response:+The [[https://iaqg.org/wp-content/uploads/2023/04/9100-2016-Series-Clarification-Table-2024-02-12.pdf|IAQG "Official" AS 9100:2016 Series Clarifications]] contains the following response:
 <blockquote>//The 9100-series clause 7.1.5.2 was not intended to force organizations to have the register specifically include the "equipment type, unique identification, location, and the calibration or verification method, frequency, and acceptance criteria.” The organization is required to have this information for equipment listed on the calibration register but not specifically in the register.//</blockquote> <blockquote>//The 9100-series clause 7.1.5.2 was not intended to force organizations to have the register specifically include the "equipment type, unique identification, location, and the calibration or verification method, frequency, and acceptance criteria.” The organization is required to have this information for equipment listed on the calibration register but not specifically in the register.//</blockquote>
  
 To further clarify, if the documented information is available, and not necessarily in the "register", [[https://iaqg.org/|IAQG]] considers the requirement to have been adequately addressed. To further clarify, if the documented information is available, and not necessarily in the "register", [[https://iaqg.org/|IAQG]] considers the requirement to have been adequately addressed.
  
-The entire [[https://www.sae.org/iaqg/projects/9100-2016_clarification_table.pdf|IAQG "Official" AS 9100:2016 Series Clarifications]] is available online.+The entire [[https://iaqg.org/wp-content/uploads/2023/04/9100-2016-Series-Clarification-Table-2024-02-12.pdf|IAQG "Official" AS 9100:2016 Series Clarifications]] is available online. 
 + 
 +To contact the SDR (Sector Document Representative) for AS 9100-series in your region, Reference: [[https://tst.iaqg.org/iaqg/publications/standards_register_tracking_matrix.pdf| Contact info. for AS Standard IDRs (IAQG Document Representatives) & SDRs (Sector Document Representatives)]]. 
 + 
 +===== 11 - Counting & Weighing Scales marked "For Reference Only" ===== 
 + 
 +__Situation:__ While auditing a Distributor, you observe their weighing scales, counting scales and length meters (for measuring the length of cable & wire), all identified as "For Reference Only"
 + 
 +The company guide explained that management had decided that these measuring devices were not critical to quality and therefore do not require calibration. 
 + 
 +(1) Is this a nonconformity? And if so, (2) is it a minor or a major nonconformity? 
 + 
 +__Answer:__ This IS a nonconformance.  
 + 
 +Every State has their own statutory requirements relating to commercial trade. In Texas, for example, the statutory requirement for measuring devices (e.g., counting scales – when used to count piece parts being sold, weighing scales – when product is sold by weight, length meters – for cable & wire sold by length) to be calibrated, is contained in:   
 + 
 +[[https://statutes.capitol.texas.gov/docs/AG/htm/AG.13.htm|TEXAS AGRICULTURE CODE]] \\ 
 +**TITLE 2. DEPARTMENT OF AGRICULTURE \\ 
 +CHAPTER 13. WEIGHTS AND MEASURES \\ 
 +SUBCHAPTER B. STANDARD WEIGHTS AND MEASURES** 
 + 
 +**Sec. 13.036.  FALSE REPRESENTATION OF COMMODITY QUANTITY.** \\  
 +A person violates this chapter if the person or the person's servant or agent: \\ 
 +(1)  sells or offers or exposes for sale a quantity of a commodity or service that is less than the quantity the person represents; 
 + 
 +And 
 + 
 +**Sec. 13.037.  USE OF INCORRECT WEIGHING OR MEASURING DEVICE.**  \\ 
 +  (a)  A person commits an offense if the person or the person's servant or agent knowingly uses an incorrect weighing or measuring device in: 
 +    (1)  buying or selling a commodity;  
 +    (2)  computing a charge for services rendered on the basis of weight or measure; or 
 +    (3)  determining the weight or measure of a commodity, if a charge is made for the determination. 
 +  (b)  For the purpose of this section, a weighing or measuring device is incorrect if it: 
 +    (1)  does not conform as closely as practicable to the official standards; 
 +    (2)  is not accurate; 
 +    (3)  is of a construction that is not reasonably permanent in adjustment or does not correctly repeat its indications; 
 +    (4)  facilitates the perpetration of fraud; or 
 +    (5)  does not conform to the specifications and tolerances established by the department under Section 13.114. 
 + 
 + 
 +While the Texas Dept. of Agriculture may seem like an odd place for such a commercial statue, this is fairly common. For example, in Texas the Texas Department of Agriculture (TDA) inspects, tests, and calibrates motor fuel metering devices (gas pumps)… and that certainly has nothing to do with agriculture. 
 + 
 +The above Texas Statute relates to all commodities being sold commercially. 
 + 
 +Where uncalibrated measurement devices are used for "//determining the weight or measure of a commodity//", then management has no way to know whether they are accurate. Therefore, this could be identified as a major nonconformity citing ISO 9001/AS9100, sec. 5.1.2 "Customer focus", which states: 
 +<blockquote>//Top management shall demonstrate leadership and commitment with respect to customer focus by ensuring that: \\ 
 +a) customer and applicable statutory and regulatory requirements are determined, understood and consistently met;//</blockquote> 
 + 
 +The nonconformity statement would read: \\ 
 +"Due to the organization not calibrating its weighing scales, counting scales and/or length meters, the Quality Management System is not entirely effective in satisfying the requirements of Texas Statute, TITLE 2, CHAPTER 13, SUBCHAPTER B, Sec. 13.037.(b)(2))."
  
-To contact the SDR (Sector Document Representative) for AS 9100-series in your region, Reference: [[https://iaqg.org/wp-content/uploads/2021/11/IAQG-Standards-Register-Tracking-Matrix-October-27-2021.pdf| Contact info. for AS Standard IDRs (IAQG Document Representatives) & SDRs (Sector Document Representatives) - Updated Oct. 2021]].