Line 41: Line 41:
 a. calibrated or verified, or both, at specified intervals, or prior to use, against measurement standards __traceable to international or national measurement standards; when no such standards exist, the basis used for calibration or verification shall be retained as documented information__;//" a. calibrated or verified, or both, at specified intervals, or prior to use, against measurement standards __traceable to international or national measurement standards; when no such standards exist, the basis used for calibration or verification shall be retained as documented information__;//"
 </WRAP> </WRAP>
-Unfortunately, the language in ISO 9001:2015 & AS9100:2016 (Rev. D) is extremely antiquated (by 55 years) in stating its requirements for "measurement traceability". In contrast, ISO 17025:2017 is much more modern and technically correct in stating:\\+Unfortunately, the language in ISO 9001:2015 & AS9100:2016 (Rev. D) is extremely antiquated in stating its requirements for "measurement traceability". In contrast, ISO 17025:2017 is much more modern and technically correct in stating:\\
 <WRAP center round box 80%> <WRAP center round box 80%>
 **ISO 17025:2017, “General requirements for the competence of testing and calibration laboratories”** \\ **ISO 17025:2017, “General requirements for the competence of testing and calibration laboratories”** \\